| General Specification |
Product name |
Automatic COD monitor |
| Model name |
CODA-500 |
| Objects |
COD concentration in water |
| Dimensions (*1) |
600(W)×510(D)×1600(H) mm |
| Mass |
Approx. 150 kg |
| Power source |
100-240V AC±10% (50/60 Hz) |
| Power consumption |
100-120 AC : Approx. 250VA 120-240 AC : Approx. 350VA |
| Installation conditions |
Indoor installation type Transient overvoltage of main power source: Overvoltage categoryII (IEC60364-44), pollution level 2 |
| Performance |
Measurement range (Upper limit of measurement is 50% of F.S.) |
0-20 mg/L 0-30,40,50,100,200,500,1000,2000 mg/L(1-dilution type) |
Repeatability (with standard solution for glucose) |
20 mg/L range |
Within ±1% F.S. (*2) |
| 30-500 mg/L range |
Within ±2% F.S. (*2) |
| 1000-2000 mg/L range |
Within ±5% F.S. |
| Stability |
Zero drift (for 24 h) |
20 mg/L range: within ± 3% F.S. (*2) 30-500 mg/L range: within ± 4% F.S. (*2) 1000-2000 mg/L range: within ± 5% F.S. |
Zero drift (for 24 h) (with standard solution for glucose) |
20 mg/L: within ± 3% F.S. (*2) 30-500 mg/L: within ± 4% F.S. (*2) 1000-2000 mg/L: within ± 5% F.S. |
| Measuring principle |
CODA-500-A (Acidic method) |
Acid potassium permanganate method at 100ºC (based on JIS K 0806) |
| CODA-500-B (Alkaline method) |
Alkaline potassium permanganate method at 100ºC |
| CODA-500-C (Acidic method) |
Acid potassium permanganate method at 100ºC (based on JIS K 0806). With cleaning function using reagent |
| Number of measurement points |
Standard: 1 point (optional: 2 points) |
| Measuring range |
Standard: 1 range (optional: 2 ranges) |
| Heating method |
Direct heating |
| End point detection |
Potentiometric titration at constant current |
| Titration method |
Micro syringe titration |
| Measurement interval |
60 minutes |
| Silver nitrate solution free method |
Available (in case of low chloride ion concentration) |
| Measuring conditions |
Ambient temperature |
2-40 ºC |
| Ambient humidity |
Relative humidity: 85% max. (without condensation) |
| Power supply voltage fluctuations |
100-240V AC ± 10% |
| Sample water conditions |
Temperature |
2-40 ºC (without freezing) |
| Flow rate (when overflow tank is used) |
2-20 L/min (when OF-5 is used), 5-20 L/min (when OF-50 is used) |
| Chloride ion concentration (for acidic method) |
CODA-500A: up to 1 times of F.S. CODA-500C: up to 100 times of F.S. (max) (for more than 100 times of F.S., select the alkaline method) |
| Sampling point |
Piping length from main unit: within 3 m |
| Blank water conditions |
Supply method |
Standard: tap water (optional: pure water tank) |
| Water quality (*3) |
Tap water without COD (hardness: 100 mg/L max.) |
| Water supply pressure |
100-500 kPa |
| Consumption |
20-420 mL (depends on measurement ranges and the setup of cleaning function) |
| Installation conditions |
Well ventilated indoor location without exposure to direct sunlight. Flat and stable location with minimized vibrations and shocks. Atmosphere free from dust, mist, corrosive gas, etc. |
| Input/output specification |
Display |
LCD color touch panel display |
| Analog output |
Number of points |
Standard: 3 points (optional: 6 points) |
| Type |
Standard: 4-20mA DC, 0-16mA DC (optional: 0-1V DC, 1-5V DC) (Default setting is 4-20mA DC. On-screen switching to 0-16mA DC is available) |
| Description |
COD concentration, time COD load, time flow rate |
| Output impedance |
900 Ω max. |
| Contact output |
Number of points |
Standard: 14 points (optional: 21 points) |
| Format |
Non voltage contact output |
| Type |
Insulated output |
| Output capacity |
250V AC, 3A / 30V DC, 3A (only 30V DC, 3A are available for "maintenance" contact. |
| Status output |
Meas., Cal., Standby, Cleaning, Blank Meas., Synchro. Idle 1, Synchro. Idle 2, Maintenance, Power, etc. |
| Warning output |
Limit warnings (COD Hi limit, Flow limit, and COD Hi Load), COD H.Hi limit, Sample Lack, Total Alarm 1-6, various gauge errors, etc. (optional: Lack Reagent) |
| Analog input |
Number of points |
Standard: 1 point (optional: 2 points) |
| Type |
Standard: 4-20mA DC (optional:1-5V DC) |
| Description |
Flow signal (full scale setting is available optionally) |
| Contact input |
Number of points |
Standard: 9 points (optional: 17 points) |
| Format |
Non voltage a contact input (open collector is available) |
| Type |
Insulated type input: common to (-) side |
| ON resistance |
100 Ω max. |
| Open voltage |
26V DC max. |
| Short-circuit current |
13 mA DC max. |
| Functions |
Meas. Start, Cal. Start, Cleaning Start, Blank meas. Start, Modify Date, Samp. Lack, Line Select, Flow Mainte, Flow Err., Flow Power OFF, and Flow No Drain |
| Load calculation |
COD load calculation |
| Memory |
1 year ( in measurement values), external USB memory. |
| Communication |
Standard: RS -485 (Optional: RS-232C) |
| Communication protocol |
Modbus ® |
| Printer |
Standard (58 mm); with automatic take-up device. |